Procedure for Freeze Drying Sediment Samples

Procedure
- place samples in open vials on the shelves of the freeze drier
- VERY CAREFULLY lift bell cover and set in place being sure not to disturb
any of your carefully placed samples
- be sure that all valves are closed (in up, that is 12:00, position
- press the condenser button and wait for about 15 minutes
- press and hold the vacuum button until you hear the vacuum motor kick on
- leave samples for at least 12 hours
- turn vacuum off by pressing and holding the vacuum button until you hear
the vacuum motor kick off
- turn off condenser by pressing that button
- very slowly release the vacuum by opening one of the valves in the bell
cover until you hear the sound of air entering the chamber
- wait for a few minutes until the full vacuum has been released
- carefully lift off bell cover and retrieve samples
- samples should still be placed in the drying oven at 50°C to ensure
complete air-dryness